Page 12
Calibrate a 2-Wire Transmitter by sourcing the input while
monitoring the output.
Works with SOURCE pH, T/C, DC V, OHMS, RTD, FREQ and
READ PRESSURE.
Move the power switch
w to READ and Double click the e
DIAL KNOB and the MAIN menu for the function in use will
appear for 15 seconds:
Turn the
e DIAL KNOB to move to the second (or third) menu
page so the word mA DISPLAY appears at the top of the
menu.
Turn the e DIAL KNOB to move through the menu. Press the e
DIAL KNOB to toggle between OFF and ON or to change the
MODE setting.
HART 250Ω - turn on the 250Ω resistor if you are powering up
a HART transmitter. This provides the loop load required for
HART communication.
EXIT MENU - exits this menu immediately and saves any
changes. Menu will automatically exit after 15 seconds of
inactivity.
mA DISPLAY
> EXIT
MODE OFF READ PWRM READ% PWRM%
HART 250Ω OFF ON
LEAK DETECT OFF ON
MAIN
> EXIT
FUNCTION T/C
UNITS °C
T/C TYPE K
COLD JUNC ON
Operating Instructions